Three‐dimensional ultrasound‐guided robotic needle placement: an experimental evaluation
Emad M. Boctor, Michael A. Choti, Everette C. Burdette, Robert J. Webster
- 发表年份
- 2008
- 引用次数
- 147
- 访问权限
- 开放获取
摘要
BACKGROUND: Clinical use of image-guided needle placement robots has lagged behind laboratory-demonstrated robotic capability. Bridging this gap requires reliable and easy-to-use robotic systems. METHODS: Our system for image-guided needle placement requires only simple, low-cost components and minimal, entirely off-line calibration. It rapidly aligns needles to planned entry paths using 3D ultrasound (US) reconstructed from freehand 2D scans. We compare system accuracy against clinical standard manual needle placement. RESULTS: The US-guided robotic system is significantly more accurate than single manual insertions. When several manual withdrawals and reinsertions are allowed, accuracy becomes equivalent. In ex vivo experiments, robotic repeatability was 1.56 mm, compared to 3.19 and 4.63 mm for two sets of manual insertions. In an in vivo experiment with heartbeat and respiratory effects, robotic system accuracy was 5.5 mm. CONCLUSIONS: A 3D US-guided robot can eliminate error bias and reduce invasiveness (the number of insertions required) compared to manual needle insertion. Remaining future challenges include target motion compensation.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991