Home /Research /Motion Analysis of a Robotic Assisted Surgery and Microsurgery System - Experimental Verification
SURGICAL

Motion Analysis of a Robotic Assisted Surgery and Microsurgery System - Experimental Verification

Mohsen Moradi Dalvand, Bijan Shirinzadeh, Yanding Qin, Willem Schwalb, Julian A. Smith, Yongmin Zhong

Year
2012
Citations
2

Abstract

Motion analysis of a parallel robot assisted minimally invasive surgery/microsurgery system (PRAMiSS) and the control structures enabling it to achieve milli/micromanipulations under the constraint of moving through a fixed penetration point or so-called remote centre-of-motion (RCM) are presented in this article. Two control algorithms are proposed suitable for minimally invasive surgery (MIS) with submillimeter accuracy and for minimally invasive micro-surgery (MIMS) with submicrometer accuracy. The RCM constraint is performed without having any mechanical constraint. Control algorithms also apply orientation constraint preventing the tip to orient relative to the soft tissues due to the robot movements. Experiments were conducted to verify accuracy and effectiveness of the proposed control algorithms for MIS and MIMS operations. The experimental results demonstrate accuracy and performance of the proposed position control algorithms.

Keywords

MicrosurgeryInvasive surgeryConstraint (computer-aided design)Motion controlComputer scienceRobotMedical roboticsPoint (geometry)SimulationArtificial intelligence

Related papers

Browse all SURGICAL papers