Self-calibration of cone-beam CT geometry using 3D-2D image registration: development and application to tasked-based imaging with a robotic C-arm

S Ouadah, J. Webster Stayman, Grace J. Gang, Ali Uneri, Tina Ehtiati, Jeffrey H. Siewerdsen

Year
2015
Citations
8

Abstract

<strong>Purpose</strong>: Robotic C-arm systems are capable of general noncircular orbits whose trajectories can be driven by the particular imaging task. However obtaining accurate calibrations for reconstruction in such geometries can be a challenging problem. This work proposes a method to perform a unique geometric calibration of an arbitrary C-arm orbit by registering 2D projections to a previously acquired 3D image to determine the transformation parameters representing the system geometry. <strong>Methods</strong>: Experiments involved a cone-beam CT (CBCT) bench system, a robotic C-arm, and three phantoms. A robust 3D-2D registration process was used to compute the 9 degree of freedom (DOF) transformation between each projection and an existing 3D image by maximizing normalized gradient information with a digitally reconstructed radiograph (DRR) of the 3D volume. The quality of the resulting “self-calibration” was evaluated in terms of the agreement with an established calibration method using a BB phantom as well as image quality in the resulting CBCT reconstruction. <strong>Results:</strong> The self-calibration yielded CBCT images without significant difference in spatial resolution from the standard (“true”) calibration methods (p-value &gt;0.05 for all three phantoms), and the differences between CBCT images reconstructed using the “self” and “true” calibration methods were on the order of 10<sup>-3</sup> mm<sup>-1</sup>. Maximum error in magnification was 3.2%, and back-projection ray placement was within 0.5 mm. <strong>Conclusion</strong>: The proposed geometric “self” calibration provides a means for 3D imaging on general noncircular orbits in CBCT systems for which a geometric calibration is either not available or not reproducible. The method forms the basis of advanced “task-based” 3D imaging methods now in development for robotic C-arms.

Keywords

Cone beam ctCalibrationComputer visionImage registrationCone beam computed tomographyArtificial intelligenceComputer scienceComputed tomographyImage (mathematics)Medicine

Related papers

Browse all OTHER papers