Home /Research /Design and development of miniature parallel robot for eye surgery
SURGICAL

Design and development of miniature parallel robot for eye surgery

Tomoya Sakai, Kanako Harada, Shinichi Tanaka, Takashi Ueta, Yasuo Noda, Naohiko Sugita, Mamoru Mitsuishi

Year
2014
Citations
17

Abstract

A five degree-of-freedom (DOF) miniature parallel robot has been developed to precisely and safely remove the thin internal limiting membrane in the eye ground during vitreoretinal surgery. A simulator has been developed to determine the design parameters of this robot. The developed robot's size is 85 mm × 100 mm × 240 mm, and its weight is 770 g. This robot incorporates an emergency instrument retraction function to quickly remove the instrument from the eye in case of sudden intraoperative complications such as bleeding. Experiments were conducted to evaluate the robot's performance in the master-slave configuration, and the results demonstrated that it had a tracing accuracy of 40.0 μm.

Keywords

RobotLimitingSimulationComputer scienceMobile robotRobot kinematicsTracingVitreoretinal surgeryArtificial intelligenceEngineering

Related papers