Evaluation of Ultra-High-Resolution Cone-Beam CT Prototype of Twin Robotic Radiography System for Cadaveric Wrist Imaging

Jan‐Peter Grunz, Andreas Max Weng, Carsten Gietzen, Maike Veyhl-Wichmann, Lenhard Pennig, Andreas Steven Kunz, Rainer Schmitt, Süleyman Ergün, Thorsten Alexander Bley, Tobias Gassenmaier

Year
2020
Citations
19

Abstract

Rationale and ObjectivesCone-beam CT (CBCT) applications possess potential for dose reduction in musculoskeletal imaging. This study evaluates the ultra-high-resolution CBCT prototype of a twin robotic X-ray system in wrist examinations compared to high-resolution multidetector CT (MDCT).Materials and MethodsSixteen wrists of body donors were examined with the CBCT scan mode and a 384 slice MDCT system. Radiation-equivalent low-dose (CTDIvol(16cm) = 3.3 mGy) and full-dose protocols (CTDIvol(16cm) = 13.8 mGy) were used for both systems. Two observers assessed image quality on a seven-point Likert scale. In addition, software-assisted quantification of signal intensity fractions in cancellous bone was performed. Fewer pixels with intermediate signal intensity were considered to indicate superior depiction of bone microarchitecture.ResultsSubjective image quality in CBCT was superior to dose equivalent MDCT with p ≤ 0.03 for full-dose and p < 0.001 for low-dose scans, respectively. Median Likert values were 7/7 (reader 1 / reader 2) in full-dose CBCT, 6/6 in full-dose MDCT, 5/6 in low-dose CBCT and 3/3 in low-dose MDCT. Intraclass correlation coefficient was 0.936 (95% confidence interval, 0.897-0.961; p < 0.001), indicating excellent reliability. Objective analysis displayed smaller fractions of “indecisive” pixels with intermediate signal intensity for full-dose CBCT (0.57 [interquartile range 0.13]) compared to full-dose MDCT (0.68 [0.21]), low-dose CBCT (0.72 [0.19]), and low-dose MDCT (0.80 [0.15]) studies. No significant difference was observed between low-dose CBCT and full-dose MDCT.ConclusionThe new CBCT prototype provides superior image quality for trabecula and bone marrow in cadaveric wrist studies and enables dose reduction up to 75% compared to high-resolution MDCT. Cone-beam CT (CBCT) applications possess potential for dose reduction in musculoskeletal imaging. This study evaluates the ultra-high-resolution CBCT prototype of a twin robotic X-ray system in wrist examinations compared to high-resolution multidetector CT (MDCT). Sixteen wrists of body donors were examined with the CBCT scan mode and a 384 slice MDCT system. Radiation-equivalent low-dose (CTDIvol(16cm) = 3.3 mGy) and full-dose protocols (CTDIvol(16cm) = 13.8 mGy) were used for both systems. Two observers assessed image quality on a seven-point Likert scale. In addition, software-assisted quantification of signal intensity fractions in cancellous bone was performed. Fewer pixels with intermediate signal intensity were considered to indicate superior depiction of bone microarchitecture. Subjective image quality in CBCT was superior to dose equivalent MDCT with p ≤ 0.03 for full-dose and p < 0.001 for low-dose scans, respectively. Median Likert values were 7/7 (reader 1 / reader 2) in full-dose CBCT, 6/6 in full-dose MDCT, 5/6 in low-dose CBCT and 3/3 in low-dose MDCT. Intraclass correlation coefficient was 0.936 (95% confidence interval, 0.897-0.961; p < 0.001), indicating excellent reliability. Objective analysis displayed smaller fractions of “indecisive” pixels with intermediate signal intensity for full-dose CBCT (0.57 [interquartile range 0.13]) compared to full-dose MDCT (0.68 [0.21]), low-dose CBCT (0.72 [0.19]), and low-dose MDCT (0.80 [0.15]) studies. No significant difference was observed between low-dose CBCT and full-dose MDCT. The new CBCT prototype provides superior image quality for trabecula and bone marrow in cadaveric wrist studies and enables dose reduction up to 75% compared to high-resolution MDCT.

Keywords

MedicineNuclear medicineCone beam computed tomographyImage qualityInterquartile rangeIsocenterRadiologyImaging phantomComputed tomographySurgery

Related papers

Browse all OTHER papers