MRI guided focused ultrasound robotic system for animal experiments

Marinos Yiannakou, Georgios Menikou, Christos Yiallouras, Cleanthis Ioannides, Christakis Damianou

Year
2017
Citations
26

Abstract

BACKGROUND: In this paper an MRI-guided focused ultrasound (MRgFUS) robotic system was developed that can be used for conducting experiments in small animals.The target for this robotic system regarding motion was to move a therapeutic ultrasound transducer in two Cartesian axes. METHODS: A single element spherically focused transducer of 3 cm diameter, focusing at 7 cm and operating at 0.4 MHz was used. The positioning device incorporates only MRI compatible materials. The propagation of ultrasound is a bottom to top approach. The 2-D positioning device is controlled by custom-made software and a custom-made electronic system which controls the two piezoelectric motors. RESULTS: The system was tested successfully in agar/silica/evaporated milk phantom for various tasks (robot motion, MR compatibility, and MR thermometry). The robotic system is capable of moving the focused ultrasound transducer to perform MR-guided focused ultrasound experiments in small animals. CONCLUSIONS: This system has the potential to be deployed as a cost effective solution for performing experiments in small animals.

Keywords

Imaging phantomUltrasoundTransducerComputer scienceUltrasonic motorUltrasonic sensorSoftwarePositioning systemBiomedical engineeringCartesian coordinate system