High resolution ion chamber array delivery quality assurance for robotic radiosurgery: Commissioning and validation
Oliver Blanck, Laura Masi, M Chan, Sebastian Adamczyk, Christian Albrecht, Marie-Christin Damme, Britta Loutfi-Krauß, Manfred Alraun, Roman Fehr, Ulla Ramm, Frank‐André Siebert, Tenzin Sonam Stelljes, Daniela Poppinga, Björn Poppe
- Year
- 2016
- Citations
- 39
- Access
- Open access
Abstract
PURPOSE: High precision radiosurgery demands comprehensive delivery-quality-assurance techniques. The use of a liquid-filled ion-chamber-array for robotic-radiosurgery delivery-quality-assurance was investigated and validated using several test scenarios and routine patient plans. METHODS AND MATERIAL: Preliminary evaluation consisted of beam profile validation and analysis of source-detector-distance and beam-incidence-angle response dependence. The delivery-quality-assurance analysis is performed in four steps: (1) Array-to-plan registration, (2) Evaluation with standard Gamma-Index criteria (local-dose-difference⩽2%, distance-to-agreement⩽2mm, pass-rate⩾90%), (3) Dose profile alignment and dose distribution shift until maximum pass-rate is found, and (4) Final evaluation with 1mm distance-to-agreement criterion. Test scenarios consisted of intended phantom misalignments, dose miscalibrations, and undelivered Monitor Units. Preliminary method validation was performed on 55 clinical plans in five institutions. RESULTS: The 1000SRS profile measurements showed sufficient agreement compared with a microDiamond detector for all collimator sizes. The relative response changes can be up to 2.2% per 10cm source-detector-distance change, but remains within 1% for the clinically relevant source-detector-distance range. Planned and measured dose under different beam-incidence-angles showed deviations below 1% for angles between 0° and 80°. Small-intended errors were detected by 1mm distance-to-agreement criterion while 2mm criteria failed to reveal some of these deviations. All analyzed delivery-quality-assurance clinical patient plans were within our tight tolerance criteria. CONCLUSION: We demonstrated that a high-resolution liquid-filled ion-chamber-array can be suitable for robotic radiosurgery delivery-quality-assurance and that small errors can be detected with tight distance-to-agreement criterion. Further improvement may come from beam specific correction for incidence angle and source-detector-distance response.
Keywords
Related papers
Campbell-Walsh urology
Alan J. Wein editor-in-chief
2012
Principles of Robot Motion: Theory, Algorithms, and Implementations
Howie Choset, Jean‐Claude Latombe
2005
Minimally Invasive versus Abdominal Radical Hysterectomy for Cervical Cancer
Pedro T. Ramírez, Michael Frumovitz, René Pareja +16 more
2018
Guideline for Management of the Clinical T1 Renal Mass
Steven C. Campbell, Andrew C. Novick, Arie S. Belldegrun +9 more
2009