The accuracy of a novel image‐guided hybrid robotic system for dental implant placement: An in vitro study
Baoxin Tao, Yuan Feng, Xingqi Fan, Kengliang Lan, Minjie Zhuang, Shigang Wang, Feng Wang, Xiaojun Chen, Yiqun Wu
- Year
- 2022
- Citations
- 44
Abstract
BACKGROUND: This in vitro study aims to evaluate the accuracy of dental implant placement by a novel image-guided hybrid robotic system for dental implant surgery (HRS-DIS). METHODS: The HRS-DIS with a 5 degree of freedom (DOF) serial manipulator and a 6 DOF Stewart platform was developed. To evaluate the accuracy of repeated drilling, the holes were prepared twice with a 2.2 mm drill. To evaluate the accuracy of dental implant placement, the entry, exit and angle deviations of dental implants were measured. RESULTS: Twenty-four holes were prepared twice, and mean (±SD) of diameters were measured as 2.2 ± 0.02 mm. A total of 160 dental implants were placed in 32 phantoms by HRS-DIS. The mean (±SD) of the entry, exit and angle deviation were 0.8 ± 0.54 mm, 0.87 ± 0.54 mm and 1.0 1 ± 0.44°, respectively. CONCLUSIONS: The results of the in vitro study preliminarily validated that the HRS-DIS could provide a high accuracy for dental implant surgery.
Keywords
Related papers
Robots and Jobs: Evidence from US Labor Markets
Daron Acemoğlu, Pascual Restrepo
2019
Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes
Darren J. Lipomi, Michael Vosgueritchian, Benjamin C. K. Tee +4 more
2011
Stretchable, Skin‐Mountable, and Wearable Strain Sensors and Their Potential Applications: A Review
Morteza Amjadi, Ki‐Uk Kyung, Inkyu Park +1 more
2016
Reach and grasp by people with tetraplegia using a neurally controlled robotic arm
Leigh R. Hochberg, Daniel Bacher, Beata Jarosiewicz +8 more
2012