首页 /研究 /Is Exactrac X-Ray System an Alternative to CBCT for Positioning Patients Head and Neck Cancers?
OTHER

Is Exactrac X-Ray System an Alternative to CBCT for Positioning Patients Head and Neck Cancers?

Stefania Clemente

发表年份
2013
引用次数
3

摘要

Purpose: To evaluate the usefulness of a 6-degrees-of freedom (6D) correction using ExacTrac Robotics system in patients with head-and-neck (HN) cancer receiving radiation therapy. Methods: Local setup accuracy was analyzed for 12 patients undergoing intensity-modulated radiation therapy (IMRT). Patient position was imaged daily upon two different protocols, cone-beam computed tomography (CBCT) and ExacTrac (ET) images correction. Setup data from either approach were compared in terms of both residual errors after correction and punctual displacement of selected regions of interest (Mandible, C2 and C6 vertebral bodies). Results: On average, both protocols achieved reasonably low residual errors after initial correction. The observed differences in shift vectors between the two protocols showed that CBCT tends to weight more C2 and C6 at the expense of the mandible, while ET tends to average more differences among the different ROIs. Conclusions: CBCT, even without 6D correction capabilities, seems preferable to ET for better consistent alignment and the capability to see soft tissues. Therefore, in our experience, CBTC represents a benchmark for positioning head and neck cancer patients.

关键词

MedicineHead and neck cancerMandible (arthropod mouthpart)Head and neckRadiation therapyNuclear medicineResidualRadiologyOrthodonticsComputer science

相关论文