Development and long‐term stability of a comprehensive daily <scp>QA</scp> program for a modern pencil beam scanning (<scp>PBS</scp>) proton therapy delivery system
Suresh Rana, Jaafar Bennouna, E. James Jebaseelan Samuel, Alonso N. Gutiérrez
- 发表年份
- 2019
- 引用次数
- 43
- 访问权限
- 开放获取
摘要
Abstract Purpose The main purpose of this study is to demonstrate the clinical implementation of a comprehensive pencil beam scanning ( PBS ) daily quality assurance ( QA ) program involving a number of novel QA devices including the Sphinx/Lynx/parallel‐plate ( PPC 05) ion chamber and HexaCheck/multiple imaging modality isocentricity ( MIMI ) imaging phantoms. Additionally, the study highlights the importance of testing the connectivity among oncology information system (OIS), beam delivery/imaging systems, and patient position system at a proton center with multi‐vendor equipment and software. Methods For dosimetry, a daily QA plan with spot map of four different energies (106, 145, 172, and 221 MeV) is delivered on the delivery system through the OIS . The delivery assesses the dose output, field homogeneity, beam coincidence, beam energy, width, distal‐fall‐off (DFO), and spot characteristics — for example, position, size, and skewness. As a part of mechanical and imaging QA , a treatment plan with the MIMI phantom serving as the patient is transferred from OIS to imaging system. The HexaCheck/ MIMI phantoms are used to assess daily laser accuracy, imaging isocenter accuracy, image registration accuracy, and six‐dimensional (6D) positional correction accuracy for the kV imaging system and robotic couch. Results The daily QA results presented herein are based on 202 daily sets of measurements over a period of 10 months. Total time to perform daily QA tasks at our center is under 30 min. The relative difference (Δ rel ) of daily measurements with respect to baseline was within ± 1% for field homogeneity, ±0.5 mm for range, width and DFO, ±1 mm for spots positions, ±10% for in‐air spot sigma, ±0.5 spot skewness, and ±1 mm for beam coincidence (except 1 case: Δ rel = 1.3 mm). The average Δ rel in dose output was −0.2% (range: −1.1% to 1.5%). For 6D IGRT QA , the average absolute difference (Δ abs ) was ≤0.6 ± 0.4 mm for translational and ≤0.5° for rotational shifts. Conclusion The use of novel QA devices such as the Sphinx in conjunction with the Lynx, PPC 05 ion chamber, HexaCheck/ MIMI phantoms, and my QA software was shown to provide a comprehensive and efficient method for performing daily QA of a number of system parameters for a modern proton PBS ‐dedicated treatment delivery unit.
关键词
相关论文
Statistical Learning Theory
Yuhai Wu, Vladimir Vapnik
1999
Artificial intelligence: a modern approach
1995
Fractional Differential Equations
Igor Podlubný
2025
Applied Nonlinear Control
Jean-Jacques Slotine, Weiping Li
1991