Kinematic control of an articulated minimally invasive surgical robotic arm
Surbhi Gupta, Sankho Turjo Sarkar, Amod Kumar
- 发表年份
- 2016
- 引用次数
- 2
摘要
The robotic arm used in minimally invasive surgery enters patient's body through a port which constrains its end-effector translation along two axes. We aim to achieve the minimally-invasive operations using a general articulated robotic arm (GARA). The algorithm is applicable to articulated robotic arm independent of its design; given only end-link is constrained. Geometric transformations based on the constraints acting on the end-link coupled with kinematic-relations obtained using conventional techniques, were used to drive a simulated 6-DOF GARA for minimally-invasive operations. The method was verified by tracing predefined planar and 3D trajectories using this simulated arm. The mean deviation of the traced trajectories was of the order of 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-03</sup> cm and the mean absolute error in maintaining remote center-of-motion (RCM) at the port was ~0 (<; 10 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">-15</sup> cm). The proposed method enabled a GARA to perform minimally-invasive operations without specialized design and with sufficient accuracy.