摘要

The maximum temperature of vertebral cancellous bone and the cortical bone is primarily influenced by the output power and the feed rate, respectively; the penetration time of cancellous bone and the cortical bone is affected by the feed rate and both of feed rate and output power, respectively. The most suitable parameters are output power of 120 W and feed rate of 2.4 mm/s for cancellous bone and output power of 120 W and feed rate of 1.6 mm/s for cortical bone.

关键词

Cancellous boneUltrasonic sensorBiomedical engineeringMaterials scienceCortical boneAnatomyMedicine