首页 /研究 /Modeling of and Experimenting with Concentric Tube Robots: Considering Clearance, Friction and Torsion
SURGICAL

Modeling of and Experimenting with Concentric Tube Robots: Considering Clearance, Friction and Torsion

Tianxiang Liu, Gang Zhang, Peng Zhang, Tianyu Cheng, Zijie Luo, Shengsong Wang, Fuxin Du

发表年份
2023
引用次数
9
访问权限
开放获取

摘要

Concentric tube robots (CTRs) are a promising prospect for minimally invasive surgery due to their inherent compliance and ability to navigate in constrained environments. Existing mechanics-based kinematic models typically neglect friction, clearance, and torsion between each pair of contacting tubes, leading to large positioning errors in medical applications. In this paper, an improved kinematic modeling method is developed. The effect of clearance on tip position during concentric tube assembly is compensated by the database method. The new kinematic model is mechanic-based, and the impact of friction moment and torsion on tubes is considered. Integrating the infinitesimal torsion of the concentric tube robots eliminates the errors caused by the interaction force between the tubes. A prototype is built, and several experiments with kinematic models are designed. The results indicate that the error of tube rotations is less than 2 mm. The maximum error of the feeding experiment does not exceed 0.4 mm. The error of the new modeling method is lower than that of the previous kinematic model. This paper has substantial implications for the high-precision and real-time control of concentric tube robots.

关键词

ConcentricKinematicsTorsion (gastropod)RobotTube (container)SimulationActuatorMechanicsClampingComputer science

相关论文

查看 SURGICAL 分类全部论文