首页 /研究 /Validation of a Novel, Fully Automated High Throughput High-Performance Liquid Chromatographic/Tandem Mass Spectrometric Method for Quantification of Pantoprazole in Human Plasma
OTHER

Validation of a Novel, Fully Automated High Throughput High-Performance Liquid Chromatographic/Tandem Mass Spectrometric Method for Quantification of Pantoprazole in Human Plasma

Yannis Dotsikas, Constantinos Apostolou, Stefanos Soumelas, Filomila Kolocouri, Afroditi Ziaka, Constantinos Kousoulos, Yannis L. Loukas

发表年份
2010
引用次数
11
访问权限
开放获取

摘要

An automated high-throughput HPLC/MS/MS method was developed for the quantitative determination of pantoprazole in human plasma. Only 100 microL plasma was placed in 2.2 mL 96 deep-well plates, and both pantoprazole and omeprazole (IS) were extracted from human plasma by liquid-liquid extraction, using diethyl ether-dichloromethane (70:30, v/v) as the organic solvent. Robotic liquid-handling workstations were used for all liquid transfer and solution preparation steps and resulted in a short sample preparation time. After vortexing, centrifugation, and freezing, the supernatant organic solvent was evaporated and reconstituted in a small volume of reconstitution solution. Sample analysis was performed by utilizing the combination of RP-HPLC/MS/MS, with positive-ion electrospray ionization and multiple reaction monitoring detection. The chromatographic run time was set at 1.8 min with a flow rate of 0.6 mL/min on a Nucleosil octylsilyl (C8) analytical column. The method was proven to be sensitive, specific, accurate, and precise for the determination of pantoprazole in human plasma. The method was applied to a bioequivalence study after per os administration of a 40 mg pantoprazole gastric retentive tablet.

关键词

ChromatographyChemistryPantoprazoleHigh-performance liquid chromatographyElectrosprayElectrospray ionizationSample preparationExtraction (chemistry)Mass spectrometryOmeprazole

相关论文

查看 OTHER 分类全部论文