首页 /研究 /Structure design for a Two-DoF myoelectric prosthetic hand to realize basic hand functions in ADLs
MANIPULATION

Structure design for a Two-DoF myoelectric prosthetic hand to realize basic hand functions in ADLs

Suguru Hoshigawa, Yinlai Jiang, Ryu Kato, Soichiro Morishita, Tatsuhiro Nakamura, Yoshiko Yabuki, Hiroshi Yokoi

发表年份
2015
引用次数
21

摘要

Prosthetic hands are desired by those who have lost a hand or both hands not only for decoration but also for the functions to help them with their activities of daily living (ADL). Prosthetic robotic hands that are developed to fully realize the function of a human hand are usually too expensive to be economically available, difficult to operate and maintain, or over heavy for longtime wearing. The aim of this study is therefore to develop a simplified prosthetic hand (sim-PH), which is to be controlled by myoelectric signals from the user, to realize the most important grasp motions in ADL by trading off the cost and performance. This paper reports the structure design of a two-DoF sim-PH with two motors to drive the CM joint of the thumb and the interlocked MP joints of the other four fingers. In order to optimize the structure, the model of the sim-PH was proposed based on which 7 sim-PHs with different structural parameters were manufactured and tested in a pick-and-place experiment. Correspondence analysis of the experimental results clarified the relationship between the hand functions and the shapes of fingers.