首页 /研究 /Effects of master-slave tool misalignment in a teleoperated surgical robot
SURGICAL

Effects of master-slave tool misalignment in a teleoperated surgical robot

Lawrence H. Kim, Clifford Bargar, Yuhang Che, Allison M. Okamura

发表年份
2015
引用次数
26

摘要

In a teleoperated system, misalignment between the master and slave manipulators can result from clutching, errors in the kinematic model, and/or sensor errors. This study examines the effects of type and magnitude of misalignment on the performance of the teleoperator. We first characterized the magnitude and direction of orientation misalignment created when clutching and unclutching during use of two surgical robots: the Raven II and the da Vinci Research Kit. We then purposely generated typical misalignments in order to measure the impact of such misalignment on user performance of a peg transfer task with the Raven II. Users were able to compensate for misalignment angles up to approximately 20 degrees in both tool orientation and camera viewpoint misalignment. These results can be used to guide the design and control of teleoperated systems for a variety of applications.

关键词

TeleoperationRobotKinematicsOrientation (vector space)Computer scienceSimulationTeleroboticsMaster/slaveTask (project management)Engineering

相关论文

查看 SURGICAL 分类全部论文