A manipulator system for 14‐gauge large core breast biopsies inside a high‐field whole‐body MR scanner

Stefan O.R. Pfleiderer, Jürgen R. Reichenbach, Tarek Azhari, Christiane Marx, Ansgar Malich, Achim Schneider, Jörg Vagner, Harald Fischer, Werner A. Kaiser

发表年份
2003
引用次数
39
访问权限
开放获取

摘要

PURPOSE: To investigate a robotic manipulator system for MR-guided breast biopsies inside a 1.5 T whole-body magnet. MATERIALS AND METHODS: Fourteen patients with suspicious breast lesions (diameter 18.6 +/- 12 mm) underwent biopsy. Patients with malignant findings underwent surgery afterwards and the histologic findings between biopsy and surgery were correlated. RESULTS: In five patients biopsy findings were histopathologically confirmed following open surgery. One tubular carcinoma was missed; one invasive cancer was underestimated. Seven patients with benign findings are still in the follow-up period. CONCLUSION: The study demonstrates the feasibility to perform breast biopsies inside the magnet of a whole-body MR scanner by using a manipulator system.

关键词

MedicineBiopsyBreast cancerScannerCore biopsyRadiologyBreast carcinomaBreast biopsyCancerMammography