Commissioning and acceptance testing of a CyberKnife linear accelerator

Subhash C. Sharma, J. Ott, J. Williams, Danny Dickow

发表年份
2007
引用次数
39

摘要

Acceptance testing and commissioning of a CyberKnife robotic stereotactic radiosurgery system was performed in April 2006. The CyberKnife linear accelerator produces a photon beam of 6 MV nominal energy, without the use of a flattening filter. Clinically measured tissue-phantom ratios, off-center ratios, and output factors are presented and compared with similar data from other CyberKnife sites throughout the United States. In general, these values agreed to within 2%.

关键词

CyberknifeLinear particle acceleratorImaging phantomRadiosurgeryNuclear medicineMedical physicsMedicineBeam (structure)PhysicsRadiation therapy