Quantitative assessment of manual and robotic microcannulation for eye surgery using new eye model

Shinichi Tanaka, Kanako Harada, Yoshiki Ida, Kyohei Tomita, Ippei Kato, Fumihito Arai, Takashi Ueta, Yasuo Noda, Naohiko Sugita, Mamoru Mitsuishi

发表年份
2014
引用次数
41

摘要

BACKGROUND: Microcannulation, a surgical procedure for the eye that requires drug injection into a 60-90 µm retinal vein, is difficult to perform manually. Robotic assistance has been proposed; however, its effectiveness in comparison to manual operation has not been quantified. METHODS: An eye model has been developed to quantify the performance of manual and robotic microcannulation. The eye model, which is implemented with a force sensor and microchannels, also simulates the mechanical constraints of the instrument's movement. Ten subjects performed microcannulation using the model, with and without robotic assistance. RESULTS: The results showed that the robotic assistance was useful for motion stability when the drug was injected, whereas its positioning accuracy offered no advantage. CONCLUSIONS: An eye model was used to quantitatively assess the robotic microcannulation performance in comparison to manual operation. This approach could be valid for a better evaluation of surgical robotic assistance.