Measuring Intracellular Calcium Fluxes in High Throughput Mode

Chris Chambers, Fiona Smith, Christine A. Williams, Sandra Marcos, Zhen Han Liu, Paul M. Hayter, Giuseppe Ciaramella, Wilma Keighley, Philip Gribbon, Andreas Sewing

发表年份
2003
引用次数
58

摘要

The measurement of intracellular calcium fluxes in real time is widely applied within the pharmaceutical industry to measure the activation of G-protein coupled receptors (GPCRhyp;s), either for pharmacological characterisation or to screen for new surrogate ligands. Initially restricted to G(q) coupled GPCRs, the introduction of promiscuous and chimeric G-proteins has further widened the application of these assays. The development of new calcium sensitive dyes and assays has provided sensitive, homogeneous assays which can be readily applied to high throughput screening (HTS). In this paper we describe the full automation of this assay type using a fluorometric imaging plate reader (FLIPR ) integrated into a Beckman/Sagian system to establish a simple robotic system that is well suited for the current medium throughput screening in this area of lead discovery. Using a recently completed HTS we discuss important determinants for FLIPR based screening, highlight some limitations of the current approach, and look at the requirements for future automated systems capable of keeping up with expanding compound files.

关键词

ThroughputHigh-throughput screeningComputer scienceG protein-coupled receptorHomogeneousDrug discoveryAutomationComputational biologyBiochemical engineeringChemistry