The geometric accuracy of frameless stereotactic radiosurgery using a 6D robotic couch system

Toru Takakura, Takashi Mizowaki, Manabu Nakata, Shinsuke Yano, Takahiro Fujimoto, Yuki Miyabe, Mitsuhiro Nakamura, Masahiro Hiraoka

发表年份
2009
引用次数
119

摘要

The aim of this paper is to assess the overall geometric accuracy of the Novalis system using the Robotic Tilt Module in terms of the uncertainty in frameless stereotactic radiotherapy. We analyzed the following three metrics: (1) the correction accuracy of the robotic couch, (2) the uncertainty of the isocenter position with gantry and couch rotation, and (3) the shift in position between the isocenter and central point detected with the ExacTrac x-ray system. Based on the concept of uncertainty, the overall accuracy was calculated from these values. The accuracy in positional correction with the robotic couch was 0.07 +/- 0.22 mm, the positional shift of the isocenter associated with gantry rotation was 0.35 mm, the positional shift of the isocenter associated with couch rotation was 0.38 mm and the difference in position between the isocenter and the ExacTrac x-ray system was 0.30 mm. The accuracy of intracranial stereotactic radiosurgery with the Novalis system in our clinic was 0.31 +/- 0.77 mm. The overall geometric accuracy based on the concept of uncertainty was 0.31 +/- 0.77 mm, which is within the tolerance given in the American Association of Physicists in Medicine report no. 54.

关键词

IsocenterRadiosurgeryRotation (mathematics)Position (finance)Computer scienceImage-guided radiation therapyNuclear medicineArtificial intelligenceMedical imagingComputer vision

相关论文

查看 SURGICAL 分类全部论文