The application accuracy of the NeuroMate robot?A quantitative comparison with frameless and frame-based surgical localization systems

Qing Hang Li, L. Zamorano, Abhilash Pandya, R. Pérez, Jianxing Gong, Fernando Díaz

发表年份
2002
引用次数
178
访问权限
开放获取

摘要

The NeuroMatetrade mark robot system (Integrated Surgical Systems, Davis, CA) is a commercially available, image-guided, robotic-assisted system used for stereotactic procedures in neurosurgery. In this article, we present a quantitative comparison of the application accuracy of the NeuroMate with that of standard frame-based and frameless stereotactic techniques. The article discusses a five-way application accuracy comparison study. The variables of our comparison and their mean errors are as follows: (1) with the robot in a frame-based configuration, the RMS was 0.86 +/- 0.32 mm; (2) with the robot in the frameless configuration, the RMS was 1.95 +/- 0.44 mm; (3) in a standard stereotactic (ZD) frame-based approach, the RMS was 1.17 +/- 0.25 mm; (4) with an infrared tracking system using the frame for fiducial registration, the RMS was 1.47 +/- 0.45 mm; (5) with an infrared tracking system using screw markers for registration, the RMS was 0.68 +/- 0.26 mm. The study was performed with 2-mm sections of CT scans. These results show that the application accuracy of the frame-based NeuroMate robot is comparable to that of standard localizing systems, whether they are frame-based or infrared tracked.

关键词

Fiducial markerFrame (networking)Computer visionComputer scienceArtificial intelligenceReference frameRobotTracking systemTracking (education)Filter (signal processing)

相关论文

查看 SURGICAL 分类全部论文